2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 424
CO- 10.0.41 PREPARE TO ANALYZE PRE AND POST CONDUCTIVITY STABILITY
a) Install a ‘dummy chamber’ in the Level Detector module and RESET ALL alarms!
b) Watch the screen for thirty (30) seconds. If a “Filling Program” banner EVER occurs, allow thirty (30)
seconds before continuing!
NOTE! Multiple “Filling Programsoccurrences indicate a hydraulic air leak that
MUST be located and
repaired or else OLC tests cancel.
c) If screen says “Standby For Test” OR the screen says “Test:…” allow the tests to finish THEN allow
three (3) minutes BEFORE continuing to procedure number CO- 10.0.42 (page 424).
CO- 10.0.42 ISOLATE FLOW PUMP DIGITAL CONTROL (FLWP)
a) Call debug screen 1. Press the keyboard’s “1” keypad for five (5) seconds.
b) Allow forty-five (45) seconds BEFORE continuing to part c.
c) Call debug screen 0. Without resetting alarms, ENSURING Flow Error
REMAINS = 0 ALSO watch FLWP for two (2) minutes. TWO (2)
possible scenarios:
1) IF (and ONLY if) FLWP continually changes more than
+/- 2 (two): Proceed to
page 52, procedure number F- 3.8.0.
2) IF FLWP does NOT change more than +/- 2 (two): Reset ALL
alarms THEN see procedure number CO- 10.0.43 (page 424).
CO- 10.0.43 PRE CONDUCTIVITY STABILITY
a) Call debug screen 5. Pre Dev val (right column) is the conductivity difference
between now and one (1) minute ago from Conductivity Cell #7. If no alarms
it changes continually!
b) After no more than two (2) minutes, if not already, Pre Dev val should fall to less than 135. TWO (2)
possible scenarios:
1) IF (and ONLY if) Pre Dev val is OR falls to less than 135: See procedure number
CO- 10.0.44 (page 425).
2) IF Pre Dev val NEVER falls below 135: Pre Cond is unstable! ENSURING no alarms occurred
see parts a AND b below:
a) Per the Figure next page, using a flashlight and WITHOUT LOOKING AWAY, for one (1)
minute, if air bubbles are flowing through Pre Dialyzer Cond Cell #7’s tubing locate and repair
the air leak!
b) If (and ONLY if) absolutely sure of no air, proceed to
page 429, procedure number
CO- 10.0.60.

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals