2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 24
F- 1.0.0 NO WATERNEVER APPEARS / (OPTIONAL) BIBAG EQUIPPED?
Figure right, TWO (2) possible scenarios 1) or 2) below:
1) IF equipped with a bibag Connector: See
procedure number F- 1.0.1 (page 24).
2) IF (and ONLY if) NO bibag Connector:
Proceed to page 25, procedure number
F- 1.0.2.
F- 1.0.1 EQUIPPED WITH BIBAG CONNECTOR
A) Connect the Bicarb (blue) connector FIRMLY to a JUG
1
of BICARB! The Bicarb connector CANNOT
be in its rinse port!
1
A bibag disposable (pouch) CANNOT be attached while troubleshooting!
B) Call debug screen 2. BICOUT MUST = 1!
C) ENSURE the Bicarb jug is FULL. If it ever runs empty Loading Pressure becomes unstable!
D) Call debug screen 0. Figure below, ENSURE the bibag Connector door is shown FULLY closed!
E) See procedure number F- 1.0.2 (page 25).
LEFT BLANK INTENTIONALLY