2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 362
CO- 1.0.8.5 AFTER RINSE CONDUCTIVITY = 10.0 MS
a) Press the [Dialysate Flow] window and set it to 800 ml/min.
b) Press ‘Enter’.
c) Allow six (6) minutes BEFORE proceeding to
page 365, procedure number CO- 2.0.0.
LEFT BLANK INTENTIONALLY