2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 277
T- 1.7.72 TEMP = 4.0 OR MORE
a) Turn the Heater Switch OFF!
b) Return NTC #2 to the
1
st
distribution board position from the left, “CON-NTC”.
c) Temperature may continue to increase for a few minutes! Allow the external meter to fall below 35.0°
C BEFORE continuing to procedure number T- 1.7.8 (page 277).
T- 1.7.8 VERIFY TEMPERATURE
a) Turn the Heater Switch ON!
b) Return to Dialysis Program (“Select Program” → ‘Dialysis’ → ‘Enter)
c) From the Home screen, set [Dialysate Flow] to 800 ml/min and press ‘Enter.
d) Allow five (5) minutes BEFORE continuing to part e!
e) Based on the Home screen’s [Temperature] reading. TWO (2) possible scenarios:
1) IF LESS THAN 35.1° C: See procedure number T- 1.8.0 (page 278).
2) IF BETWEEN 35.1 and 38.9° C: A Temperature problem is not present at THIS time. To check
possible intermittent temperature problems proceed to page 240, procedure number T- 1.0.8.
LEFT BLANK INTENTIONALLY

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals