2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 194
CLEAN- 4.3.0 VALVE #29’S ‘DOT’ ALWAYS BLUE
Call debug screen 1 (Figure right). Does VERR
(lower right column) = 0?
Yes VERR = 0! Proceed to
page
195, procedure number
CLEAN- 4.3.2.
No VERR = 1 OR more! See procedure
number CLEAN- 4.3.1 (page 194).
CLEAN- 4.3.1 VERR = 1 OR MORE
a) Call debug screen 0
(Figure right).
b) Ignoring the top Flow
Error window, in part c,
you will watch the
2
nd
window down, Valve
Error.
c) Watching for one (1) minute, ignoring a ‘blink to 1’ that lasts less than one (1) second, does Valve
Error EVER = 1 for LONGER THAN two (2) seconds?
Yes Valve Error = 1 LONGER THAN two (2) seconds! Proceed to page 207, procedure
number CLEAN- 7.0.0.
No Valve Error = 0 OR ‘blinks to 1’ for less than one (1) second! Close the shunt door THEN
proceed to page 224, TROUBLESHOOTING VALVE ERRORS IN CLEANING
PROGRAMS
.
LEFT BLANK INTENTIONALLY

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals