2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 635
DIASAFE- 7.0.0 VALVE #28 OKAY
a) Return FILTER #92 to Valve #28!
b)
Per the Figure below, ENSURE ‘TUBE A’ goes to VALVE #24 AND ‘TUBE B’ goes to VALVE
#26!
c) Are you ABSOLUTELY SURE the DiaSafe
®
tubing is plumbed correctly?
Yes Tubing is plumbed properly! See procedure number DIASAFE- 8.0.0 (page 636).
No Tubing is backwards! Per the Figure, attach the tubing properly and repeat the test.