2008T Troubleshooting Guide
P/N 490292 Rev. A
F- 2.6.0 MOTORS OKAY / ISOLATE ‘OPTIONAL’ DIALYSATE SAMPLER
a) The shunt door MUST remain open till instructed!
b) Press the keyboard’s ‘Esc’ key.
c) At the bottom of the screen, press the Test & Options tab. DO NOT start the tests!
d) Per the Figure below, does the OPTIONAL [Dialysate Sampler] button appear?
Yes The [Dialysate Sampler] button should NOT appear! Refer to the NOTE below!
No The [Dialysate Sampler] does NOT appear! See procedure number F- 2.7.0 (page 41).
Test & Options Screen / Optional Dialysate Sampler button
F- 2.7.0 [DIALYSATE SAMPLER] DOES NOT APPEAR
Call the Home screen. Figure right, if the TMP window is RED a TMP alarm
is present. TWO (2) possible scenarios:
1) IF (and ONLY if) the TMP window is WHITE: See procedure number F- 2.8.0 (page 42).
2) IF the TMP window is RED: Perform parts A THROUGH D below:
A) Press and release the ‘Reset’ key then immediately press and hold it for three (3) seconds.
B) Allow thirty (30) seconds before going to part C.
C) If a TMP alarm reoccurs perform parts A and B up to twice more.
D) Allow twenty (20) seconds then see procedure number F- 2.8.0 (page 42).
The [Dialysate Sampler] button
appears if the Dialysate Sampler
option has been accidentily
turned on. FIGURE3CThis may
cause intermittent flow errors
and/or tubing to blow off. This
option MUST be set to “No” from
Service Mode → Options →