2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 590
d) IMPORTANT! Allow fifteen (15) seconds then watch the tubing opening for up to five (5) minutes. Is
there flow or dripping (possibly slowly) from it?
Yes Flow or dripping seen! See procedure number OVER- 4.0.1 (page 590).
No No flow OR dripping seen! Proceed to
page 591, procedure number OVER- 5.0.0.
OVER- 4.0.1 FLOW OR DRIPPING
a) Figure right, unplug Valve #41 from
distribution board position labeled
V27, IN-V”.
b) Does flow or dripping continue from the Air
Gap tubing?
Yes Flow or dripping continues! Valve #41 is leaking and must be replaced! To LOCATE Valve
#41 refer to Figure 35 (page 211).
No Flow or dripping stops! Perform parts a
THROUGH c below:
a) Figure right, unplug the Float’s connector from
distribution board position X5, FLOAT-SW.
b) Check inside the vacant “X5” position for
corrosion or damage. Damage indicates
the distribution board needs to be
replaced!
c) Watch for up to five (5) minute. Does flow or dripping continue?
Yes Flow or dripping continues! The Actuator-Test Board is bad. To located the
board refer to Figure 4A (page 10).
No Flow or dripping stops! Float Switch #5 is bad! To LOCATE Float Switch #5 refer
to Figure 28 (page 140).
LEFT BLANK INTENTIONALLY

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals