2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 39
Yes ! EMPTY = 1: See procedure number F- 2.3.0 (page 39).
No ! EMPTY = 0: Either the arterial (red) dialyzer line IS NOT connected to the shunt properly OR
FOUR (4) possible bad components: 1) Actuator-Test Board
a
OR; 2) Functional Board
a, b
;
3) Arterial dialyzer line shunt door switch OR; 4) Motherboard
a
a
To LOCATE the boards, refer to Figure 4A (page 10).
b
To prevent “Cond Offset Failure” place the machine into T and C Mode (refer to
OPERATING MODES (page 19))
F- 2.3.0 ! EMPTY = 1 / BC VALVES NOT CYCLING
a) To ENSURE the shunt door is OPEN, CVRCLS (2
nd
column from left) = 0 always!
b) Watch DIALVLO (2
nd
column from left) for thirty (30) seconds.
TWO (2) possible scenarios:
1) IF (and ONLY if) DIAVLO REMAINS ALWAYS = 0! See procedure number F- 2.4.0 (page 39).
2) IF DIALVLO cycles between 0 and 1 (possibly rapidly) OR is remaining = 1! Valve #24 may be
in a malfunctioning electrical state! Proceed to page 711, Section 26.
F- 2.4.0 VERR = 0 / BC VALVES NOT CYCLING
Call debug screen 1. Watch FLWP (upper right), for thirty (30) seconds! TWO (2)
possible scenarios:
1) IF (and ONLY if) FLWP REMAINS LESS THAN 255: See procedure number F- 2.5.0 (page 40).
2) IF FLWP is ALWAYS = 255! The machine may NOT be going in Dialysis Program! Perform parts a
THROUGH c below:
a)
Turn the machine OFF!
b) Turn the machine on and return to Dialysis Program (“Select Program” ‘Dialysis’ ‘Enter)!
c) Call debug screen 1. ENSURING a “No Water” NEVER occurs watch FLWP for one (1) FULL
minute! TWO (2) possible scenarios:
1) IF (and ONLY if) FLWP IS NOT REMAINING = 255! See procedure number F- 2.5.0
(page 40).
2) IF FLWP is REMAINING ALWAYS = 255 (i.e. NEVER less than 255): With the machine off,
swap in the following components (Component List below) one at a time then, in between,
repeat parts b and c to test each one in between until FLWP does NOT REMAIN 255.
Component List: 1) Actuator-Test Board
a
; 2) Functional Board
a, b
; 3) Motherboard
a
.
a
To LOCATE the boards, refer to Figure 4A (page 10).
b
To prevent “Cond Offset Failure”
place the machine into T and C Mode (refer to OPERATING MODES, page 19))

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals