2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 168
1) IF (and ONLY if) Valve Error = 0 OR ‘blinks to 1’ for LESS THAN two (2) seconds: See
procedure number CLEAN- 1.0.4 (page 168)
2) IF (and ONLY if) Valve Error = 1 MORE THAN (2) seconds: Proceed to
page 207, procedure
number CLEAN- 7.0.0.
CLEAN- 1.0.4 VALVE ERROR = 0 / BIBAG EQUIPPED?
Figure right, TWO (2) possible scenarios:
1) IF (and ONLY if) NOT equipped with a
bibag Connector: Turn the machine OFF
then proceed to
page 607, procedure
number LEAKING- 4.0.0.
2) IF equipped with a bibag Connector: See
parts a AND b below:
a) Call debug screen 14 (Figure below).
b) Locate Val1 Err AND Val2 Err. TWO (2) possible scenarios:
1) IF (and ONLY if) BOTH Val1 Err AND Val2 Err = 0: Turn the machine OFF then proceed to
page 607, procedure number LEAKING- 4.0.0.
2) IF Val1 Err AND/OR Val2 Err = 1: Proceed to
page 727. SECTION 29 – BIBAG: VALVE 1
OR 2 ERROR.

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals