2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 36
F- 1.0.9 LOADING PRESSURE MORE THAN 29 PSI
Adjust Loading Pressure Valve #65’s* nut counterclockwise but no more than a few turns. Can you adjust
PEAK pressure to between 23 and 25 psi (Yes or No)? * To LOCATE Valve #65 refer to Figure 6 (page 22)
Yes Loading Pressure between 23 and 25 psi! Call
debug screen 0. WITHOUT LOOKING AWAY,
watch Flow Error for five (5) minutes OR until if it
EVER = 1 indicating a Flow problem. TWO (2) possible scenarios:
1) IF (and ONLY if) Flow Error EVER = 1, even just once: Return to (ABOVE) procedure
number F- 1.0.5 (page 31).
2) IF Flow Error REMAINS = 0: A flow problem is NOT occurring at this time but other
problems may be! Refer to the Table of Contents for whatever other problem may be
occurring.
No Loading Pressure REMAINS more than 25 psi! A) Turn the machine OFF; B) Proceed to
page
613, procedure number LEAKING- 6.0.0.
LEFT BLANK INTENTIONALLY

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals