2008T Troubleshooting Guide
P/N 490292 Rev. A
Yes ‘Bob’ moving! See part E.
No ‘Bob’ NOT moving! ENSURING the machine was in RINSE, return to Dialysis Program
THEN proceed to page 23, SECTION 1 - FLOW ERRORS IN DIALYSIS PROGRAM
E) Watch for ten (10) minutes ENSURING a “No Water” or a Flow Error NEVER occurs.
F) See procedure number CO- 1.0.8.4 (page 361).
CO- 1.0.8.4 AFTER 10 MINUTE RINSE / ISOLATE COND CELL #7
a) Connect to acid and bicarb.
b) Return to Dialysis Program (Select Program” → ‘Dialysis’ → ‘Enter’)!
c)
IMMEDIATELY call the Home screen. Is [Conductivity] = 10 mS?
Yes [Conductivity] = 10.0 mS! See procedure number CO- 1.0.8.5 (page 362).
No [Conductivity] = MORE THAN 10.0 mS! See parts a THROUGH c below:
a) Not likely! BE VERY SURE the machine
was in RINSE for ten (10) FULL minutes!
b) Figure right, inside the distribution board,
unplug the 5
th
CONNECTOR CAP from the
LEFT, 6
th
position from the LEFT i.e.,
position “COND”!
c) Is [Conductivity] = 10 mS now?
Yes [Conductivity] = 10.0 mS! Pre-Dialyzer Conductivity Cell # 7* is bad. *To LOCATE
Cell #7, refer to the Figure next page).
No [Conductivity] = more than 10.0 mS! See parts a AND b below
a) Not likely! BE VERY SURE you unplugged Conductivity Cell #7’s connector from
position, “X7, COND”! If [Conductivity] is STILL more than 10.0 mS continue to
part b.
b) Figure below, TWO (2) possible bad components: 1) Bad Sensor Board cable OR
2) Bad distribution board.