2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 361
Yes ‘Bob’ moving! See part E.
No Bob’ NOT moving! ENSURING the machine was in RINSE, return to Dialysis Program
THEN proceed to page 23, SECTION 1 - FLOW ERRORS IN DIALYSIS PROGRAM
E) Watch for ten (10) minutes ENSURING a “No Water” or a Flow Error NEVER occurs.
F) See procedure number CO- 1.0.8.4 (page 361).
CO- 1.0.8.4 AFTER 10 MINUTE RINSE / ISOLATE COND CELL #7
a) Connect to acid and bicarb.
b) Return to Dialysis Program (Select Program ‘Dialysis’ → ‘Enter’)!
c)
IMMEDIATELY call the Home screen. Is [Conductivity] = 10 mS?
Yes [Conductivity] = 10.0 mS! See procedure number CO- 1.0.8.5 (page 362).
No [Conductivity] = MORE THAN 10.0 mS! See parts a THROUGH c below:
a) Not likely! BE VERY SURE the machine
was in RINSE for ten (10) FULL minutes!
b) Figure right, inside the distribution board,
unplug the 5
th
CONNECTOR CAP from the
LEFT, 6
th
position from the LEFT i.e.,
position “COND”!
c) Is [Conductivity] = 10 mS now?
Yes [Conductivity] = 10.0 mS! Pre-Dialyzer Conductivity Cell # 7* is bad. *To LOCATE
Cell #7, refer to the Figure next page).
No [Conductivity] = more than 10.0 mS! See parts a AND b below
a) Not likely! BE VERY SURE you unplugged Conductivity Cell #7’s connector from
position, “X7, COND”! If [Conductivity] is STILL more than 10.0 mS continue to
part b.
b) Figure below, TWO (2) possible bad components: 1) Bad Sensor Board cable OR
2) Bad distribution board.

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals