EasyManuals Logo
Home>Fresenius Medical Care>Medical Equipment>2008T

Fresenius Medical Care 2008T Troubleshoot Manual

Fresenius Medical Care 2008T
768 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #31 background imageLoading...
Page #31 background image
2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 27
F- 1.0.3 CHAMBER #69 SAYSNO AIR” / TROUBLESHOOT FILLING PROGRAM FAILURE
a) To avoid damage turn the machine
OFF!
b) Figure right, TWO (2) checks:
Check #1: At the top of the distribution board,
ENSURE the Sensor Cable is
plugged in SECURELY!
Check #2: ENSURE a resistor plug HAS NOT
been prematurely plugged into the
5
th
position from the left!
c) Turn the machine on and return to Dialysis
Program (“Select Program ‘Dialysis’ ‘Enter’)!
d) Figure right, from here forward, if (and ONLY if) a “Dial Valve
Failure” OR “Act Byp Valve Fail” banner EVER appears
proceed to page 711, Section 26
e) Call debug screen 0. Based on Chamber #69’s
text box now, TWO (2) possible scenarios:
1) IF it says “Air” always: Return to ABOVE
procedure number
F- 1.0.2.11 (page 26).
2) IF it continues to EVER say “No Air”: Filling Program is not working! See parts a AND b below:
a) Before performing part b NOTE this page number THEN perform INITIAL CHECKS (page 7).
b) FIVE (5) possible bad components: 1) CBE board
1
; 2) Sensor Board
2
; 3) Functional Board
3
;
4) Distribution board; 5) Motherboard.
1
A) Swap in a known good CBE board; B) With the Air Sensor’s connector REMAINING
UNPLUGGED return to Dialysis Program (“Select Program” → ‘Dialysis’ → ’Enter’);
C) From debug screen 0, if Chamber #69’s text box now says “Air” the previous CBE board
is bad.
2
A) With the machine off, swap in a known good Sensor Board; B) Place the machine into T
and C Mode (refer to OPERATING MODES (page 19); C) With the Air Sensor’s connector
REMAINING UNPLUGGED, return to Dialysis Program; D) From debug screen 0, if
Chamber #69’s text box now says “Air” the previous Sensor Board is bad.
3
A) With the machine off, swap in a known good Functional Board; B) Pace the new
Functional Board into T and C Mode; C) With the Air Sensor’s connector REMAINING
UNPLUGGED, return to Dialysis Program; D) From debug screen 0, if Chamber #69’s text
box now says “Air” the previous Functional Board is bad.

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals