EasyManuals Logo
Home>Fresenius Medical Care>Medical Equipment>2008T

Fresenius Medical Care 2008T Troubleshoot Manual

Fresenius Medical Care 2008T
768 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #339 background imageLoading...
Page #339 background image
2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 335
SECTION 5 - CONDUCTIVITY / ON LINE CLEARANCE PROBLEMS
External leaks, “No Water” and Flow Errors cause Conductivity problems. If either occur at ANY TIME
address them first! To avoid error observe ALL stated times below!
A) From the Home screen, ENSURE [Dialysate Flow] is on* and has been set to 800 ml/min for five (5)
minutes!
* If [Dialysate Flow] is blinking flow is off! Flow MUST be on!
B) If [Conductivity] is less than 12.6 mS, Figure right, ENSURE BOTH
connector plug O-rings are present! If not this may be the problem!
C) If the Automated Tests are running (screen reads “Test:.…”) allow them to finish.
D)
Remove the ‘dummy chamber’ from the Level Detector.
E) Press the [Temperature] window. Is Temp Setting
ALREADY = 37.0° C?
Yes “Temp Setting” already = 37.0° C! See part F.
No Adjust “Temp Setting” to 37.0° C THEN continue to part F.
F) Press ‘Enter’ to return to [Temperature].
G)
DO NOT RESET ALARMS!
H) Call debug screen 0. If Flow Error EVER = 1, even just
once, there is a Flow Error! WITHOUT LOOKING AWAY
watch it for, TWO (2) possible scenarios:
Scenario #1: FOUR (4) minutes, if it was necessary to adjust “Temp Setting” to 37.0° C in part E.
Scenario #2: TWO (2) minutes, if it WAS NOT necessary to adjust “Temp Setting” to 37° C in part E.
I) Is Flow Error is ALWAYS = 0?
Yes Flow Error ALWAYS = 0! See part J.
No Flow Error = 1 even if just once! Proceed to page 23, SECTION 1 - FLOW ERRORS IN
DIALYSIS PROGRAM
J) Using a flashlight, ENSURE large air bubbles are NOT moving into the machine
through the acid and bicarb inlet tubing! If air is seen check the connector(s).
K) Call the Home screen. WITHOUT LOOKING AWAY, watch [Temperature]
AND [Conductivity] for one (1) minute. Based first on
[Temperature],
TWO (2) possible scenarios next page:

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals