2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 86
F- 7.4.0 VERIFY DRAIN FLOW
a) The “Select Program” banner remains up till instructed otherwise!
b) Was there ANY drain flow at all?
Yes There was some flow! Proceed to page 87, procedure number F- 7.4.3
No There was no flow at all! See procedure number F- 7.4.2 (page 86).
F- 7.4.2 NO DRAIN FLOW
Figure below, EXCEPT for valves #38 and #31, ENSURE are ALL OTHER valves, especially Valve #30,
are plugged PROPERLY into their distribution board positions?
Yes Valves plugged in correctly! See procedure number F- 7.4.3 (page 87).
No You located a valve plugged wrong! Return ALL VALVES to their distribution board PROPER
positions and return to
page 84, procedure number F- 7.3.2.
LEFT BLANK INTENTIONALLY