EasyManuals Logo
Home>Fresenius Medical Care>Medical Equipment>2008T

Fresenius Medical Care 2008T Troubleshoot Manual

Fresenius Medical Care 2008T
768 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #151 background imageLoading...
Page #151 background image
2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 147
MOTORS- 4.6.0 ISOLATE MOTOR VOLTAGE
a) Figure right, avoiding the VACANT positions on the left
and right, return the motor connector,
without the
cap, to its distribution board position.
b) ENSURE the connector is aligned properly!
c) Set your CALIBRATED volt meter to DC voltage (V
DC
).
d) Measure, inside the distribution board, between the
terminals where the wires are soldered. IMPORTANT! ENSURE good contact with BOTH
terminals!
e) Per Table 5 below, perform the indicated Response based on measured voltage.
Table 5 – Motor Troubleshooting
MOTOR
DISTRIBUTION
BOARD
POSITION
DC VOLTAGE
READING
Your RESPONSE
DEAERATION P20, “DEGAS-P”
15 volts DC or more
TWO (2)
possible bad components:
1) Bad deaeration motor OR
2) Intermittent 24V Power Harness
connection. See Figure below.
Less than 12 volts DC
Bad motor voltage! See procedure number
MOTORS- 4.7.0 (page 148)
FLOW P21, “FLOW-P”
5 volts DC or more
TWO (2)
possible bad components:
1) Bad flow motor OR
2) Intermittent 24V Power Harness
connection. See Figure below.
Less than 5 volts DC
Bad motor voltage! See procedure number
MOTORS- 4.7.0 (page 148)
24V Power Harness (Behind the card cage / Inside the Power Supply)

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals