2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 125
F- 15.2.0 DEAP LESS THAN OR = 75 / ISOLATE DEAERATION MOTOR
Per the Figure below, is the DEAERATION MOTOR shaft rotating COUNTERCLOCKWISE (CCW)?
Yes Rotating CCW! See procedure number F- 15.2.1 (page 125).
No IS NOT rotating CCW! NOTE
ONLY the DEAERATION MOTOR will be checked then proceed to
page 141, TROUBLESHOOTING MOTORS.
Figure 26 – Deaeration Motor Rotating CCW
F- 15.2.1 DEAERATION MOTOR ROTATING CCW
Using the handle end* of a screwdriver, push HARD on and release the DEAERATION MOTOR’S shaft
several times. Can you make it stop rotating and REMAIN stopped?
Yes If SURE the motor stops! TWO (2) possible bad components: 1) Deaeration Motor (possibly
brushes) OR; 2) Deaeration Pump head.
No Motor continues to rotate! See procedure number F- 15.2.2 (page 126).