2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 177
CLEAN- 1.5.1 ISOLATE POSSIBLE AIR LOCK
a) CAUTION! Fluid and tubing may be HOT!
b) Figure right, remove the
clear tubing from the
Deaeration Pump’s INPUT nozzle. Keep the
tubing at pump level to encourage gravity
flow from it!
c) The tubing is from hydrochamber D. If any flow from it
direct it into the bucket for TWO (2) minutes BEFORE
continuing to part d.
d) Now direct the tubing into the cylinder for ONE (1) minute! More than three hundred (300) ml collected?
Yes More than 300 ml! See procedure number CLEAN- 1.5.2 (page 177).
No Less than 300 ml! Leaving the tubing off and the pump unplugged, proceed to
page 183,
procedure number CLEAN- 2.0.0.
CLEAN- 1.5.2 MORE THAN 300 ML / BIBAG?
a) Reattach the Deaeration Pump’s
tubing!
b) Figure right, TWO (2) possible scenarios:
1) IF equipped with a bibag Connector:
See procedure number CLEAN- 1.5.3
(page 178).
2) IF NOT equipped with a bibag Connector:
Proceed to page 182, procedure number
CLEAN- 1.6.0.
LEFT BLANK INTENTIONALLY

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals