2008T Troubleshooting Guide
P/N 490292 Rev. A
P- 2.0.1 SYSTEM INITIALIZATION DOES NOT REACH 100%
a) To prevent damage, turn the machine OFF!
b) Reconnect the ACTUATOR cable!
c) Figure below, unplug ALL MODULES ONLY including the Blood Pressure module! DO NOT
unplug the 24V Harness or shunt!
d) From here forward, ignore “Art BP No Comm” OR “Blood Still Sensed!” if they occur!
e) Turn the machine ON (fan running). Does System Initialization
reach 100%?
Yes System Initialization reaches 100%! See ISOLATING THE MODULES this page.
No System Initialization does NOT reach 100%! See procedure number P- 2.0.2 (page 699).
ISOLATING THE MODULES
A) Turn the machine OFF!
B) Your choice, plug one of the modules back in TO WHERE IT BELONGS*.
* CAUTION! Damage may occur if any module is reverse connected with another. NEVER plug
the Blood Pressure module in with the machine on!
C) Turn the machine on (fan running).
D) Repeat parts A through D until System Initialization DOES NOT reach 100% indicating the last module
plugged in (or its cable) is causing the problem. If after plugging in ALL modules and System Initialization
reaches 100% either one of the modules was originally plugged in wrong OR the problem may be heat
related and requires running the machine in Heat Disinfect to locate it.