2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 469
FIL- 3.1.0 PRESSURE TEST HYDRAULICS (1)
a) Figure right, connect the Four-Way Assembly
(P/N 150034) to the dialyzer connectors.
b) Place the Four-Way in the dialyzer holderl!
ENSURE a transducer protector IS NOT in the
‘to syringe’ tubing segment!
c) If using a NEO-2 attach to the +Port (top, red port).
If using a 90XL attach to the Pressure Module’s
Gauge Port.
d) DO NOT allow tension or kinks in the Four-Way
tubing segments!
e)
CLOSE THE SHUNT DOOR!
f) Call debug screen 2. To ENSURE the shunt door
is closed CVRCLS (2
nd
column from left) = 1!
g) PULL on the syringe plunger. Can you achieve between
-245 and -255 mmHg on the external meter?
Yes -250 mmHg achieved. Clamp the Four Ways’s ‘to syringe’ tubing segment to hold the
pressure then see procedure number FIL- 3.2.0 (page 469).
No -250 mmHg COULD NOT be achieved! ENSURE the transducer protector, at the meter, is
not wet OR consider replacing it! If OKAY, see procedure number FIL- 3.3.0X (page 469).
FIL- 3.2.0 -250 ACHIEVED / PRESSURE TEST HYDRAULICS (2)
Watch the meter for one (1) minute. Does pressure HOLD +/- 15 mmHg for one (1) minute?
Yes Pressure holds! Proceed to page 470, procedure number FIL- 3.5.0.
No Pressure DOES NOT hold! See procedure number FIL- 3.3.0X (page 469).
FIL- 3.3.0X PRESSURE NOT ACHIEVED OR DID NOT HOLD / ISOLATE FOUR-WAY
a) Figure right, clamp BOTH Four-Way Dialyzer Line
tubing segments.
b) Can you achieve -250 mmHg and HOLD it (+/- 15
mmHg for one minute) now?
Yes -250 achieved and HOLDS!
IMPORTANT! Remove BOTH clamps from both Four-Way
Dialysate Line segments and proceed to
page 509, procedure number TMP- 4.0.0
No Either the transducer protector at the meter is wet OR a Four-Way tubing connection is leaking.
Locate and repair the problem then return to (ABOVE) procedure number FIL- 3.0.0 (
page
468).

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals