EasyManuals Logo
Home>Fresenius Medical Care>Medical Equipment>2008T

Fresenius Medical Care 2008T Troubleshoot Manual

Fresenius Medical Care 2008T
768 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #490 background imageLoading...
Page #490 background image
2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 486
PDIA OR ADIA is IS NOT between 3.0 and 6.0 OR is NOT stable continued:
b) Continuous flow, more than 0.2 ml per minute, from the connector?
Yes Continuous flow! A) Return the connector to the shunt and close the door;
B) Turn the machine OFF; C) Proceed to
page 572, SECTION 18A –
DIAGNOSTIC VALVE LEAK TESTS.
No flow! See parts a THROUGH c below:
a)
Return the connector to the shunt AND close door!
b) Call debug screen 0 to ENSURE the TOP balancing chamber valves
are REMAINING BLUE!
c) Figure right, watch Valve #41’s ‘dot’ for one (1) minute.
It should stay white! TWO (2) possible scenarios:
1) IF (and ONLY if) Valve #41’s ‘dot’ STAYS white! Proceed
to
page 505, procedure number TMP- 3.0.0.
2) IF Valve #41’s ‘dot’ cycles between white and blue: A) Turn
the machine OFF! B) Proceed to
page 572, SECTION 18A –
DIAGNOSTIC VALVE LEAK TESTS.
TMP- 1.0.2 PDIA AND ADIA STABLE BETWEEN 3.0 AND 7.0
a) At the bottom of the screen, press the ‘Dialysate’ tab.
b) Press ‘Enter’.
c) Figure right, if (and ONLY if) the screen’s ‘status’ bar
is NOT BLANK see procedure number TMP- 1.0.3
(page 487). If BLANK repeat parts a and b!
LEFT BLANK INTENTIONALLY

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals