EasyManuals Logo
Home>Fresenius Medical Care>Medical Equipment>2008T

Fresenius Medical Care 2008T Troubleshoot Manual

Fresenius Medical Care 2008T
768 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #30 background imageLoading...
Page #30 background image
2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 26
Chamber #69 says “No Air” OR toggles continued:
a) Figure right, inside the distribution board,
ENSURE the Air Sensor’s connector is
UNPLUGGED from 5
th
position FROM THE
LEFT!
b) TWO (2) possible scenarios:
1) IF (and ONLY if) Chamber #69 now says
“Air” always: See procedure number
F- 1.0.2.11 (page 26).
2) IF Chamber #69 continues to say “No Air” or toggles between “Air” and “No Air”:
Proceed to
page 27, procedure number F- 1.0.3.
F- 1.0.2.11 AIR ALWAYS
a) Figure right, from here forward, if (and ONLY if) a “Dial
Valve Failure” OR “Act Byp Valve Fail” banner EVER
appears proceed to page 711, Section 26.
b) Call debug screen 1.
c) Allow two (2) minutes OR until if FILACT (middle column) = 1?
Yes FILACT = 1! Proceed to page 28, procedure number F- 1.0.4.
No AFTER two (2) minutes FILACT REMAINS = 0! Look at VERR (lower right).
TWO (2) possible scenarios:
1) IF (and ONLY if) VERR = 0: See procedure number F- 1.0.3 (page 27).
2) IF VERR = 1 or more: Proceed to
page 38, procedure number F- 2.0.0.
LEFT BLANK INTENTIONALLY

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals