2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 597
OVER- 8.2.0 MAXIMUM PRESSURE IS BETWEEN 18 AND 20 PSI / ISOLATE PRESSURE REGULATOR
This procedure checks Pressure Regulator #61’s ability to maintain pressure over time.
a) IMPORTANT! Turn the machine OFF!
b) Leave the water ON!
c) NOTE gauge pressure then allow fifteen (15) minutes before continuing!
d) Does pressure increase more than 2 psi over what was noted in part c?
Yes Pressure increased more than 2 psi! TWO (2) possibilities: 1) Incoming water pressure is
more than 105 psi OR; 2) Bad Regulator #61. Regulator #61 can be rebuilt (Rebuild kit,
P/N 190934).
No Pressure did NOT increase over time. Pressure regulator #61 is okay. The overflow problem
should be solved!