2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 288
T- 2.1.4 DIODE 17 CHECKED GOOD OR POWER CONTROL BOARD WAS REPLACED
a) IMPORTANT! Figure below, ENSURE all cables are PROPERLY connected to the Power Control
board!
NOTE! A good diode 17 does NOT necessarily mean the Power Control board is good
b) If troubleshooting a DIALYSIS PROGRAM temperature problem return to it (“Select Program” →
‘Dialysis’ → ‘Enter’). If a HEAT DISINFECT problem return to HEAT DISINFECT.
c)
Set to volt meter to AC voltage (~ V, V
AC
).
d) Figure right, measure BETWEEN the Heater
Connector’s BROWN and BLUE wires. TWO (2)
possible scenarios:
1) IF more than 100.0 volts AC! Return NTC #2’s
connector to the
1st distribution board position
from the left, “CON-NTC”. The new boards fixed
the problem!
2) IF less than 100.0 volts AC! See procedure
number T- 2.1.5 (page 288).
T- 2.1.5 HEATER VOLTAGE LESS THAN 10 VOLTS AC
a) Set your meter to DC voltage (V
DC
).
b) Connect the black lead to chassis ground.
c) Measure again at the Power Logic boards
T5 TOP pin. If necessary, refer to Figure 44 (page 282).
d) LESS THAN two (2.0) volts DC?
Yes T5 LESS THAN 2.0 volts! Proceed to
page 291, procedure number T- 2.1.8.
No T5 between 2.0 and 28.0 volts! See procedure number T- 2.1.6 (page 289).

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals