2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 72
F- 7.0.3 ISOLATE ACTUATOR-TEST BOARD
a) To prevent damage, turn the machine OFF!
b) Swap in a known good* Actuator-Test Board**.
* Note: Known good = has been tested in another machine that is NEVER issuing Flow Errors! If a
known good Actuator-Test board is NOT used here AND the problem is a bad board you may
perform unnecessary, and time consuming, component replacement!
** To LOCATE the Actuator-Test board refer to the Figure (previous page).
c) Close the card cage!
d) Turn the machine on return to Dialysis Program (“Select Program” Dialysis → ‘Enter’)!
e) ENSURE a “No Water” alarm NEVER occurs!
f) Call debug screen 0. WITHOUT LOOKING AWAY, watch Flow Error for up to FIVE (5) minutes OR
until it EVER = 1, even if just once. TWO (2) possible scenarios:
1) IF (and ONLY if) Flow Error EVER = 1! See procedure number F- 7.0.4 (page 73) to test for a
restricted Balancing Chamber valve!
2) IF Flow Error ALWAYS = 0! Problem solved! The previous Actuator-Test Board may be bad.
LEFT BLANK INTENTIONALLY

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals