2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 87
F- 7.4.3 ISOLATE ACTUATOR-TEST BOARD
Per the Figure below, touch valve #30’s (black) solenoid. Is it HOT?
Yes Valve #30’s solenoid is hot! See procedure number F- 7.4.4 (page 87).
No Valve #30’s solenoid is NOT hot! See parts a THROUGH c below:
a) Return valve #38 AND #31 to their distribution board positions.
b) Return the Deaeration AND Flow Pump to their distribution board positions.
c) NOTE ONLY VALVE #30 will be checked and proceed to page 210,
TROUBLESHOOTING A VALVE.
Figure 19 – Valve #30
F- 7.4.4 ISOLATE VALVE #30
a) Per the Figure above, check through valve #30’s INPUT and OUTPUT tubing for bio-growth restrictions.
b) Remove valve #30’s INPUT tubing and attach a syringe, with tubing attached that will fit snug over the
valve’s nozzle.
c) Push or pull on the syringe plunger. If valve #30 is open the plunger moves with no resistance. TWO (2)
possible scenarios next page:

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals