2008T Troubleshooting Guide
P/N 490292 Rev. A
Figure 57 – Hydraulics Rear View
CO- 3.4.0 [CONDUCTIVITY] WITHIN 0.4 OF TCD / VERIFY CONDUCTIVITY ALARM LIMITS
What color is the Home screen’s [Conductivity] window now? TWO (2) possible scenarios:
1) IF (and ONLY if) pale yellow / white: See procedure number CO- 3.5.0 (page 374).
2) IF RED: See parts a THROUGH e below:
a) At the bottom of the screen, press the ‘Dialysate’
tab.
b) Figure right, if necessary adjust the Limits so that
the ‘Actual’ Conductivity is CENTERED
between them.
c) Press ‘Enter’!
d) Allow up to two (2) additional minutes for the
‘Actual’ Conductivity window to turn white!
e) See procedure number CO- 3.5.0 (page 374).
CO- 3.5.0 VERIFY MEASURED CONDUCTIVITY
a) Allow up to two (2) additional minutes OR until the external flow indicator’s ‘bob’ starts to move up and
down!
b) Based on the meter’s reading versus the machine’s [Conductivity] window. TWO (2) possible
scenarios:
1) IF (and ONLY if) the meter and the [Conductivity] window are within 0.15 of each other:
Proceed to
page 410, procedure number CO- 9.8.0.
2) IF the meter and the [Conductivity] window are NOT within 0.15 of each other! Proceed to
page 383, procedure number CO- 6.0.0.