2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 719
DVF- 5.0.0 PIN 43 (VALVE #38) LESS THAN 40
a) Figure below, unplug Valve #38 from the distribution board.
b) Using a flashlight, check inside the vacant distribution board position. If corrosion or damage is
located the distribution board needs to be replaced!
c)
Leave Valve #38 unplugged until instructed.
d) Turn the machine on AND return to Dialysis Program (“Select Program” →
‘Dialysis’ → ‘Enter’)!
e) From the Home screen, set [Dialysate Flow] to 500 ml/min and press ‘Enter’.
f) Call debug screen 0 then allow thirty (30) seconds.
g) Do the balancing chamber dotsbegin to cycle between blue and white?
Yes The balancing chamber ‘dots’ are cycling! Balancing Chamber Valve #38* OR its blue wire
harness is bad!
* To LOCATE Valve #38 refer to Figure 36 (page 211).
No a) Return Valve #38 to its distribution board position
b) See (ABOVE) procedure number DVF- 1.2.0 (page 713).

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals