2008T Troubleshooting Guide
P/N 490292 Rev. A
DVF- 5.0.0 PIN 43 (VALVE #38) LESS THAN 40 Ω
a) Figure below, unplug Valve #38 from the distribution board.
b) Using a flashlight, check inside the vacant distribution board position. If corrosion or damage is
located the distribution board needs to be replaced!
c)
Leave Valve #38 unplugged until instructed.
d) Turn the machine on AND return to Dialysis Program (“Select Program” →
‘Dialysis’ → ‘Enter’)!
e) From the Home screen, set [Dialysate Flow] to 500 ml/min and press ‘Enter’.
f) Call debug screen 0 then allow thirty (30) seconds.
g) Do the balancing chamber ‘dots’ begin to cycle between blue and white?
Yes The balancing chamber ‘dots’ are cycling! Balancing Chamber Valve #38* OR its blue wire
harness is bad!
* To LOCATE Valve #38 refer to Figure 36 (page 211).
No a) Return Valve #38 to its distribution board position
b) See (ABOVE) procedure number DVF- 1.2.0 (page 713).