2008T Troubleshooting Guide
P/N 490292 Rev. A
CO- 10.0.46 BOTH PRE AND POST COND ARE STABLE
a) Pre Stbl Ct (pre-conductivity stability counter) starts at 90 and decreases one unit
per second if Pre Dev val remains less than 135. If not already, after no more than
two (2) minutes it should fall to 0!
b) Post Stbl Ct (post-conductivity stability counter) starts at 70 and decreases one
unit per second if Post Dev val remains less than 70. If not already, after no more
than two (2) minutes it should fall to 0.
c) Call the Home screen and press the [Kt/V AF] button at the bottom of the screen.
d) Press the [OLC Self-Test] button then ‘Enter’.
Does the “Online Clearance Self-Test”
banner appear?
Yes The “Online Clearance Self-Test” banner appeared! IMMEDIATELY proceed to
page 428, procedure number CO- 10.0.50.
No The OLC Test did NOT start! ENSURING no alarms AND the shunt door is CLOSED, TWO
(2) possible scenarios:
1) IF “Can’t Run OLC Test” banner appeared: See procedure number CO- 10.0.48
(page 426).
2) IF “OLC steps not calculated” appeared: Allow two (2) minutes then attempt to start
the OLC Test again. TWO (2) possible scenarios i) or ii) below:
i) IF “Online Clearance Self-Test” banner appears: IMMEDIATELY proceed
to page 428, procedure number CO- 10.0.50.
ii) IF “OLC steps not calculated” banner reoccurs: Read before performing!
Proceed to
page 384, procedure number CO- 6.2.0 to calibrate COND
CELLS. If (and ONLY if) you return here (“OLC steps not calculated”), TWO
(2) possible bad components: 1) Actuator-Test Board OR; 2) Functional
Board.
CO- 10.0.48 “CAN’T RUN OLC TEST” APPEARED
a) Reset all alarms!
b) Call debug screen 5.
c) If Pre Stbl Ct AND Post Stbl Ct are BOTH = 0 OR goes to 0 within two (2) minutes proceed through
the following three (3) steps until if the OLC Test starts. If (and ONLY if) one or the other DOES NOT
go to 0 within two (2) minutes proceed to page 429, procedure number CO- 10.0.60.
Step #1: Allow two (2) minutes then attempt to start the OLC Self-Test again. If “Online Clearance
Self-Test” banner appears proceed to
page 428, procedure number CO- 10.0.50. If
“Can’t Run OLC Test” reoccurs continue to step #2 next page.