2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 427
Step # 2: Read before parts a THROUGH e BEFORE performing them:
a) Turn the machine OFF then back on.
b) Return to Dialysis Program with [Dialysate Flow] set to 500 ml/min.
c) Allow six (6) minutes ENSURING the Home screen’s [Conductivity] stabilizes
between 13.0 and 14.3 mS.
d)
Reset all alarms!
e) Read before performing! Return to page 423, procedure number CO- 10.0.40. If
(and ONLY if) you return here continue to step #3.
Step #3: Read parts a THROUGH e below BEFORE performing them!
a) Perform an Acid Clean per procedure.
b) Return to Dialysis Program with [Dialysate Flow] set to 500 ml/min.
c) Allow six (6) minutes, ENSURING the Home screen’s [Conductivity] stabilizes between
13.0 and 14.3 mS.
d)
Reset all alarms!
e) Read before performing! Return to page 423, procedure number CO- 10.0.40. If (and
ONLY if) you return here proceed to
page 384, procedure number CO- 6.2.0 to
calibrate COND CELLS. After the calibration if (and ONLY if) you return here, THREE
(3) possible bad components listed below:
Component List: 1) Power Logic Board; 2) Actuator-Test Board; 3) Functional Board.
LEFT BLANK INTENTIONALLY

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals