2008T Troubleshooting Guide
P/N 490292 Rev. A
Step # 2: Read before parts a THROUGH e BEFORE performing them:
a) Turn the machine OFF then back on.
b) Return to Dialysis Program with [Dialysate Flow] set to 500 ml/min.
c) Allow six (6) minutes ENSURING the Home screen’s [Conductivity] stabilizes
between 13.0 and 14.3 mS.
d)
Reset all alarms!
e) Read before performing! Return to page 423, procedure number CO- 10.0.40. If
(and ONLY if) you return here continue to step #3.
Step #3: Read parts a THROUGH e below BEFORE performing them!
a) Perform an Acid Clean per procedure.
b) Return to Dialysis Program with [Dialysate Flow] set to 500 ml/min.
c) Allow six (6) minutes, ENSURING the Home screen’s [Conductivity] stabilizes between
13.0 and 14.3 mS.
d)
Reset all alarms!
e) Read before performing! Return to page 423, procedure number CO- 10.0.40. If (and
ONLY if) you return here proceed to
page 384, procedure number CO- 6.2.0 to
calibrate COND CELLS. After the calibration if (and ONLY if) you return here, THREE
(3) possible bad components listed below:
Component List: 1) Power Logic Board; 2) Actuator-Test Board; 3) Functional Board.
LEFT BLANK INTENTIONALLY