2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 71
F- 7.0.2 LOADING PRESSURE ‘OKAY(2)
A) CAUTION! To prevent damage, turn the machine OFF!
B) To avoid pulling cables loose, GENTLY open the card cage.
C) Per the Figure below, THREE (3) checks:
CHECK #1 CAREFULLY check the ENTIRE LENGTH of the Actuator Cable for damage!
CHECK #2 ENSURE the Actuator Cable is plugged in securely at both ends!
CHECK #3 ENSURE the valves are plugged into their CORRECT distribution board positions!
D) Was a problem located during the above three (3) checks?
Yes ENSURING the problem was repaired, see procedure number F- 7.0.2.1 (page 71).
No No problem found! See procedure number F- 7.0.3 (
page 72).
F- 7.0.2.1 PROBLEM WITH ACTUATOR CABLE OR VALVE NOT PLUGGED IN WAS LOCATED
a) Close the card cage!
b) Turn the machine on and return to Dialysis Program (“Select Program” → ‘Dialysis’ → ‘Enter’)!
c) Allow five (5) minutes then set [Dialysate Flow] to 800 ml/min and press ‘Enter.
d) Call debug screen 0, WITHOUT LOOKING AWAY, watch Flow Error for up to FIVE (5) minutes OR
until it EVER = 1, even if just once! TWO (2) possible scenarios:
1) IF (and ONLY if) Flow Error EVER = 1! See procedure number F- 7.0.3 (page 72).
2) IF Flow Error ALWAYS = 0! A flow error is not presenting at this time!

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals