2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 438
CO- 14.0.00 V104 Er C = 1
NOTE! These procedures assume the filter, inside the Bicarb connector, is clean!
a) Turn the machine OFF!
b)
Per the Figures below, FOUR (4) checks for tubing kinks and / or restrictions between:
Check #1 The Bicarb (blue) concentrate connector and the TOP, INPUT side of Valve #104.
Check #2 Valve #104’s BOTTOM, OUTPUT side and the second from the bottom port of the bibag Air
Separator Chamber
Check #3 The very bottom port of bibag Air Separator Chamber and Conductivity Cell #113.
Check #4 Conductivity Cell #113 and the Bicarb Pump’s input BOTTOM, WHITE nozzle.
c) If no restrictions were located, see
procedure number CO- 14.0.01 (page 440).