EasyManuals Logo
Home>Fresenius Medical Care>Medical Equipment>2008T

Fresenius Medical Care 2008T Troubleshoot Manual

Fresenius Medical Care 2008T
768 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #186 background imageLoading...
Page #186 background image
2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 182
CLEAN- 1.6.0 MACHINE NOT BIBAG
EQUIPPED
a) Figure right, return the Deaeration Pump to
distribution board position, “P20”!
b)
ENSURE THE DEAERATION MOTOR IS
RUNNING!
c) Figure right, turn Valve #65’s nut clockwise
(tighter) attempting to achieve a PEAK Loading
Pressure of between 23 and 25 psi?
Yes Between 23 and 25 psi! Proceed to
page 189, procedure number
CLEAN- 2.4.0.
No Less than 22 psi! See parts a AND b
below:
a) ENSURE more than 300 ml per minute
input flow to the Deaeration pump i.e.
no recurring air lock!
b) Swap the following components (see
Component List below) one at a time,
with known good then, in between
attemp to adjust Loading Pressure.
When a PEAK pressure of more than 23
psi can be achieved the last component
swapped in is the problem.
COMPONENT LIST: 1) Deaeration Pump head; 2) Deaeration Motor; 3) Valve #65*;
4) Hydrochamber.
* Figure right, ensure threads are visible under
Valve #65’s nut. If no threads are visible either the
wrong spring is installed** under Valve #65’s nut
OR Valve #65 is bad!
* * Refer to Figure 6 (page 22).

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals