2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 73
F- 7.0.4 FLOW ERROR = 1 / ISOLATE ‘FRESH SIDEFLOW PATHS
a) Plug the red concentrate connector into its rinse
port to call the “Select Program” screen but DO
NOT press any keys.
b) Balancing Chamber valve connectors will be switched to test fresh side’ flow paths. Flow is measured
into a 1000 ml GRADUATED cylinder for one (1) minute. Good flow = more than 800 ml every minute!
c) ENSURING “Select Program” REMAINS up, perform PARTS 1 AND 2 EVEN IF PART 1 yields bad
flow!
PART 1: a) Remove the RED DIALYZER connector from the shunt door and place it into a bucket!
b) Plug valve #35’s connector into valve #30’s distribution board position, “V30”. Leave
valve #30 unplugged!
c) Hold the connector over the bucket and watch flow (if any) for one (1) minute. If good
initially but significantly slows indicates an intermittent bad valve #31 or #35. If no flow,
ENSURE valve #35’s connector is placed properly in position “V30”. If okay see part d!
d) Measure flow, from the RED connector, into the cylinder for one (1) minute.
e) Return valve #35’s connector to its distribution board position!
f) Record Part 1’s measurement
THEN perform PART 2!
PART 2: a) Return the dialyzer connector to the bucket.
b)
Plug valve #37’s connector into valve #30’s position, “V30”.
c) Watch flow (if any) from the connector for one (1) minute. If good initially but
significantly slows indicates a bad valve #33 or #37. If no flow, ENSURE valve #37’s
connector is placed properly in position “V30”. If okay see part d.
d) Measure flow, from the RED connector, into the cylinder, for one (1) minute!
e) Return the connector to the shunt door.
f) Leaving valve #37 in valve #30’s position for now, analyze BOTH Part 1 AND Part 2’s
measurements per Table 3 below:
Table 3 – Parts 1 AND 2
PART 1’s
measurement
PART 2’s
measurement
Your Response
Less than 800 ml Less than 800 ml
See procedure number F- 7.0.5 (page 79)
Less than 800 ml More than 800 ml
See procedure number F- 7.3.33 (page 75)
More than 800 ml Less than 800 ml
See procedure number F- 7.4.44 (page 77)
More than 800 ml More than 800 ml
See procedure number F- 7.3.1 (page 83)

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals