2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 66
F- 6.2.0 TROUBLESHOOT ‘CONTINUOUS’ DRAIN FLOW
Drain flow NEVER stops! From the Home screen, is [Conductivity] more than 15.0 mS?
Yes [Conductivity] more than 15.0! ENSURING Loading Pressure (Rinse port gauge) is
peaking to between 23 and 27 psi, see procedure number F- 6.3.0 (page 67).
No [Conductivity] remaining less 15.0! This procedure checks if Valve #43 is sticking open.
See parts a AND b below:
a)
Per the Figure below, double clamp valve #43’s OUTPUT tubing. NOTE! The
output tubing extends towards the front of the machine!
b) Recheck drain flow. TWO (2) possible scenarios:
1) IF (and ONLY if) drain flow becomes Pulse → Stop → Pulse → Stop
:
Remove the clamps from valve #43. TWO (2) possible bad components: 1) Bad
Actuator-Test* Board OR; 2) Bad valve #43.
*
To LOCATE the Actuator-Test board, refer to Figure 4A (page 10)
2) IF drain flow continues to be ‘continuous’: Remove the clamps from valve
#43 then proceed to page 69, procedure number F- 6.5.0.
Figure 11 – Valve #43 Location

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals