2008T Troubleshooting Guide
P/N 490292 Rev. A
P- 2.0.0 INITIALIZATION DOES NOT COMPLETE / ISOLATE PUMPS
a) To prevent damage, turn the machine OFF!
b) Figure below unplug the ACTUATOR CABLE from the distribution board.
c) Turn the machine on (fan running). Does System Initialization
reach 100%?
Yes System Initialization reaches 100%! See ISOLATING THE PUMPS this page.
No System Initialization does NOT reach 100%! See procedure number P- 2.0.1 (page 698).
ISOLATING THE PUMPS
Possible 24 volt ‘short circuit’ at the Acid, Bicarb, Heparin OR UF Pump.
A) Turn the machine OFF!
B)
Reconnect the ACTUATOR cable!
C) Figure below, from the distribution board, unplug one of the pumps (your choice) then continue to part C
to see if this pump is the problem.
D) Turn the machine on (fan running). Does System Initialization reaches 100%?
Yes System Initialization reaches 100%! The unplugged pump may be causing the problem! To
confirm this turn the machine off, plug the pump back in, and turn the machine on. If System
Initialization does NOT reach 100% the pump is bad.
No System Initialization does NOT reach 100%! Turn the machine off and repeat parts B AND C until
System Initialization DOES reach 100%. If after ALL pumps have all been unplugged and System
Initialization still does NOT reach 100% the Actuator Cable OR the distribution board may be bad.