EasyManuals Logo
Home>Fresenius Medical Care>Medical Equipment>2008T

Fresenius Medical Care 2008T Troubleshoot Manual

Fresenius Medical Care 2008T
768 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #248 background imageLoading...
Page #248 background image
2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 244
Yes [Conductivity] window is WHITE! See procedure number T- 1.0.133 (page 244).
No [Conductivity] window is RED Allow up to three (3) minutes OR until if the [Conductivity]
window turns WHITE. TWO (2) possible scenarios below:
1) IF the window turns WHITE: See procedure number T- 1.0.133 (page 244).
2) IF after three (3) minutes the window remains RED: Proceed to
page 320,
procedure number T- 7.0.0.
T- 1.0.133 VERIFY ‘OUT OF BYPASS’
Call debug screen 0. Figure right, is Valve #24’s ‘dot’ blue OR white?
1) IF (and ONLY if) BLUE: See procedure number T- 1.0.134
(page 244).
2) IF WHITE: See parts a THROUGH c below:
a) Call debug screen 2. To ENSURE
the shunt door is CLOSED,
CVRCLS (2
nd
column from left) =
1!
b) From the Home screen, ENSURE the [Conductivity] AND [Temperature] windows are pale
yellow/white!
c) Call debug screen 0. Watch Valve #24’s ‘dot’ for up to (2) more minutes OR until if it turns blue. TWO
(2) possible scenarios:
1) IF (and ONLY if) the ‘dot’ turns blue: See procedure number T- 1.0.134 (page 244).
2) IF the ‘dot’ remains white AND the [Temperature] window is white: Proceed to page 320,
procedure number T- 7.0.0.
T- 1.0.134 VALVE #24’S ‘DOT’ IS BLUE / VERIFY ‘OUT OF BYPASS’ CIRCUIT
Call the Home screen. ENSURING [Dialysate Flow] is set to 500 ml/min, is the external flow indicator’s ‘bob’
rising at least ¼ way up in the sight tube?
Yes ‘Bob’ moving! Proceed to
page 250, procedure number T- 1.0.20.
No ‘Bob’ NOT moving! See parts a THROUGH c below:
a) ENSURE the shunt door is CLOSED!
b) ENSURE the [Temperature] AND [Conductivity] windows are NOT RED!
c) If ‘bob’ still is not moving AND Valve #24’s ‘dot’ is blue there may be masked Flow Error.

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals