EasyManuals Logo
Home>Fresenius Medical Care>Medical Equipment>2008T

Fresenius Medical Care 2008T Troubleshoot Manual

Fresenius Medical Care 2008T
768 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #367 background imageLoading...
Page #367 background image
2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 363
CO- 1.0.9 VERIFY DIALYSATE LIMITS
a) Press the ‘Dialysate’ tab (Figure below).
Dialysate SCREEN
b) TCD is based on Dialysate Composition and indicates what the ‘Actual’ Conductivity would be if
everything was perfect.
c) Using [Alarm Width], spread Conductivity Limits as wide as they will go and press ‘Enter’.
d) Using [Alarm Position], set the upper Limit to 0.5 above TCD and press ‘Enter’.
e) Based on the ‘Actual’ [Conductivity] window (Figure right),
TWO (2) possible scenarios:
1) IF (and ONLY if) [Conductivity] is STABLE between
13.4 and 14.3 mS: See procedure number CO- 1.0.10
(page 364).
2) ALL OTHERS: Proceed to
page 365, procedure number
CO- 2.0.0.

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals