2008T Troubleshooting Guide
P/N 490292 Rev. A
PHT- 2.0.3 FILACT REMAINS = 0
TWO (2) possible scenarios:
1) IF (and ONLY if) “Get Neg TMP” is NOT presenting! See procedure number PHT- 3.0.0 (page 523).
2) IF “Get Neg TMP” is still presenting! See parts a AND b below:
a) Figure right, remove the Fluid Sample Connector from its
port and hold it UP as shown!
b) Watch for TWO (2) minutes or until you see at least one
strong pulse that squirts into the room more than six (6)
feet (Yes or No)?
Yes Strong pulse(s) seen! TWO (2) possible scenarios:
1) IF (and ONLY if) the “Get Neg TMP” banner is still presenting: Proceed to
page
483, SECTION 9 – TMP PROBLEMS.
2) ALL OTHER banners: See procedure number PHT- 3.0.0 (page 523).
No Strong pulse(s) NEVER seen! Is the “Get Neg TMP” banner still presenting?
Yes “Get Neg TMP” still presenting! Proceed to
page 548, SECTION 14 – UF PUMP
PROBLEMS
No See procedure number PHT- 3.0.0 (page 523).