2008T Troubleshooting Guide
P/N 490292 Rev. A
F- 1.0.7 LOADING PRESSURE IS NOT ‘OKAY’ / BIBAG EQUIPPED?
Figure right, TWO (2) possible scenarios:
1) IF (and ONLY if) NOT equipped with a
bibag Connector: See procedure number
F- 1.0.8 (page 34).
2) IF equipped with a bibag Connector:
See parts a AND b below:
a) Call debug screen 0.
b) Figure right, does Valve #104’s symbol
appear?
Yes Valve #104 appears! Refer to
TABLE 1 below and respond
as instructed.
No Valve #104 DOES NOT appear!
THREE (3) possible problems:
1) Loose bibag Interface Board*
ribbon cable OR; 2) Bad bibag
Interface Board OR; 3) Bad Actuator-Test board.
* To LOCATE the Interface Board refer to Figure 4C (page 11)
Table 1 – DEBUG SCREEN 0 (VALVE #104)
Is Valve #104’s ‘dot’ BLUE?
Your Response
Yes (blue) See procedure number F- 1.0.8 (page 34)
No (white)
Proceed to
page 37, procedure number F- 1.0.10.