2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 33
F- 1.0.7 LOADING PRESSURE IS NOT ‘OKAY’ / BIBAG EQUIPPED?
Figure right, TWO (2) possible scenarios:
1) IF (and ONLY if) NOT equipped with a
bibag Connector: See procedure number
F- 1.0.8 (page 34).
2) IF equipped with a bibag Connector:
See parts a AND b below:
a) Call debug screen 0.
b) Figure right, does Valve #104’s symbol
appear?
Yes Valve #104 appears! Refer to
TABLE 1 below and respond
as instructed.
No Valve #104 DOES NOT appear!
THREE (3) possible problems:
1) Loose bibag Interface Board*
ribbon cable OR; 2) Bad bibag
Interface Board OR; 3) Bad Actuator-Test board.
* To LOCATE the Interface Board refer to Figure 4C (page 11)
Table 1 – DEBUG SCREEN 0 (VALVE #104)
Is Valve #104’s ‘dot’ BLUE?
Your Response
Yes (blue) See procedure number F- 1.0.8 (page 34)
No (white)
Proceed to
page 37, procedure number F- 1.0.10.

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals