2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 376
CO- 4.2.0 ISOLATE CONCENTRATE INLET SYSTEMS (2).
a) Return the acid and bicarb CONNECTORS HARD into their rinse ports.
b) Place into RINSE.
c) ENSURING no external leaks, see procedure number CO- 4.2.1 (page 376).
CO- 4.2.1 ISOLATE PUMP FEEDBACK
a) Call debug screen 1. Watch NO EOS AND ALWEOS (third column from right)
for one (1) FULL minute.
b) Proceed per Table 8 below:
Table 8 – EOS
NO EOS ALWEOS YOUR RESPONSE
0 (always) 0 (always)
See procedure number CO- 4.2.2 (page 376).
Does not matter
1 (EVER)
Proceed to
page 441, SECTION 6 - CONCENTRATE
PUMP ERRORS.
1 (EVER)
Does not matter
Proceed to
page 441, SECTION 6 - CONCENTRATE
PUMP ERRORS.
CO- 4.2.2 ISOLATE POTENTIAL LEAKS
a) Figure right, ENSURE NO external LEAKS
at the Acid AND Bicarb Pumps.
b) Plug into acid and bicarb!
c) Return to Dialysis Program (“Select
Program” → ‘Dialysis’ → ‘Enter’)!
d) Allow forty-five (45) seconds BEFORE
continuing to part e.
e) Is air / concentrate now drawn, TOWARDS
THE MACHINE, in ONE direction ONLY?
Yes Air / concentrate now drawn into the machine in ONE DIRECTION ONLY! See procedure
number CO- 4.2.3 (page 377).
No Air / concentrate NOT seen OR moving back and forth! Proceed to
page 378, procedure
number CO- 4.4.0.

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals