2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 69
F- 6.5.0 TMP ALARM?
a) IMPORTANT! From the Home screen, set [Dialysate Flow] to 500 ml/min and pressEnter’!
b) TWO (2) possible TMP window scenarios:
1) IF the TMP window is NOT red: See procedure number F- 6.5.2 (page 69).
2) IF the TMP window is RED (TMP alarm): A) Press and release the ‘Reset’ key then press and
hold it for three (3) seconds; B) Allow thirty (30) seconds; C) If a TMP alarm reoccurs attempt
RESET up to twice more BEFORE continuing to procedure number F- 6.5.2 (page 69).
F- 6.5.2 ISOLATE TMP STABILITY
TMP is STABLE if the TMP window remains white AND does NOT change more than 60 in three (3)
minutes. TWO (2) possible scenarios:
1) IF (and ONLY if) TMP is UNSTABLE: See procedure number F- 6.6.0 (page 69).
2) IF TMP is STABLE: A procedure, in a different Section, is performed next
. NOTE this page and
procedure number (F- 6.5.2) as you may be prompted to return to here. Perform parts A through C
below:
A) BEFORE continuing to part B, proceed to
page 580, to perform SECTION 19 - TESTING FOR
A LEAKING BALANCING CHAMBER DIAPHRAGM.
B) If a balancing chamber diaphragm leak was not located in part A, return to Dialysis Program
(“Select Program→ ‘Dialysis’ → Enter’)!
C) Proceed to
page 70, procedure number F- 7.0.0.
F- 6.6.0 TMP IS UNSTABLE
a) A procedure, in a different Section, is performed next. NOTE this page and procedure number (F- 6.6.0)
as you may prompted to return to here.
b) BEFORE continuing to part c, proceed to
page 572, to perform SECTION 18A DIAGNOSTIC
VALVE LEAK TESTS
c) If a leaking balancing chamber valve was not located in part b, return to Dialysis Program (“Select
Program→ ‘Dialysis’ → Enter)!
d) From the Home screen, set [Dialysate Flow] to 800 ml/min and pressEnter’.
e) Call debug screen 6. WITHOUT LOOKING AWAY, watch BC Switch for up to five (5) minutes.
ENSURING a “No Water” alarm NEVER occurs does BC Switch EVER = 897 or more even just once?
Yes BC Switch = 897 or more (constant or once in a while): Proceed to page 70, procedure
number F- 7.0.0.
No BC Switch NEVER, EVER = 897 or more: If TMP remains unstable proceed to
page 483,
SECTION 9- TMP PROBLEMS . If TMP is stable the flow problem is no longer presenting.

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals