2008T Troubleshooting Guide
P/N 490292 Rev. A
F- 14.0.8 VALVE #26 OKAY
a) Remove the clamp from Valve #26!
b) Obtain a bucket!
c)
Per the Figure below, remove the tubing from Valve #24’s INPUT and direct into the bucket
Figure 25 – Valve #24
d) From Calibrate Hydraulics → Flow Pressure.
e) Press ‘Enter’ THEN ENSURE the Flow Motor is rotating!
f) Measure from Valve #24’s tubing for one (1) minute!
g) Press ‘Enter’ to return to “Calibrate Hydraulics”. More than 800 ml collected?
Yes More than 800 ml! See procedure number F- 14.0.81 (page 123).
No A) Reconnect Valve #24’s tubing. B) TWO (2) possible bad components: 1) Restricted
DiaSafe
©
filter OR 2) Restricted tubing inside the DiaSafe
©
housing.
F- 14.0.81 GOOD FLOW FROM VALVE #24’S / ISOLATE ACTUATOR-TEST BOARD / VALVE #24
a) Reconnect Valve #24’s tubing.
b) Per the Figure above, touch valve 24’s (black) solenoid. Is it warm (i.e. energized)?
Yes Solenoid is warm! TWO (2) possible bad components: 1) Actuator-Test Board; 2) Valve #24.
No Solenoid in NOT warm! NOTE
ONLY VALVE #24 will be checked and proceed to page
210, TROUBLESHOOTING A VALVE.