2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 123
F- 14.0.8 VALVE #26 OKAY
a) Remove the clamp from Valve #26!
b) Obtain a bucket!
c)
Per the Figure below, remove the tubing from Valve #24’s INPUT and direct into the bucket
Figure 25Valve #24
d) From Calibrate Hydraulics → Flow Pressure.
e) Press ‘Enter’ THEN ENSURE the Flow Motor is rotating!
f) Measure from Valve #24’s tubing for one (1) minute!
g) Press ‘Enter’ to return to “Calibrate Hydraulics. More than 800 ml collected?
Yes More than 800 ml! See procedure number F- 14.0.81 (page 123).
No A) Reconnect Valve #24’s tubing. B) TWO (2) possible bad components: 1) Restricted
DiaSafe
©
filter OR 2) Restricted tubing inside the DiaSafe
©
housing.
F- 14.0.81 GOOD FLOW FROM VALVE #24’S / ISOLATE ACTUATOR-TEST BOARD / VALVE #24
a) Reconnect Valve #24’s tubing.
b) Per the Figure above, touch valve 24’s (black) solenoid. Is it warm (i.e. energized)?
Yes Solenoid is warm! TWO (2) possible bad components: 1) Actuator-Test Board; 2) Valve #24.
No Solenoid in NOT warm! NOTE
ONLY VALVE #24 will be checked and proceed to page
210, TROUBLESHOOTING A VALVE.

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals