2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 468
FIL- 3.0.0 FILACT = 0 / CREATE VENOUS PRESSURE
This procedure creates Venous Pressure to ensure Valve #25 remains open for the subsequent test:
a)
From the Home screen, set [Dialysate Flow] to “OFF” and press ‘Enter’!
b) Call debug screen 0. ENSURE all eight (8) balancing chamber valves (#31 through #38) ‘dots’ are
REMAINING white (Flow is Off)!
c) Dialysate Flow MUST REMAIN OFF till instructed otherwise!
d) Figure below, attach a syringe, with a piece of tubing attached that will fit SNUG to the Level
Detector’s INNER P
ven
port.
e) Call debug screen 1.
f) Push on the syringe plunger until VEN (middle column) is between 400 and 450.
g) Clamp the syringe tubing to keep the pressure!
h) If VEN falls more than three (3) mmHg in thirty (30) seconds there is a leak at the P
ven
port OR inside
the module.
i) Leaving the clamp in place to keep VEN at more than 400, see procedure number FIL- 3.1.0 (page
469).

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals