EasyManuals Logo
Home>Fresenius Medical Care>Medical Equipment>2008T

Fresenius Medical Care 2008T Troubleshoot Manual

Fresenius Medical Care 2008T
768 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #546 background imageLoading...
Page #546 background image
2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 542
4. Per the Figure below, check the DiaSafe
®
filter’s headers. If no leaks continue to step #5.
5. Look for small leaks, from the tubing and tubing fittings, INSIDE the DiaSafe
®
housing! If no leaks
continue to step #6.
Figure 82 – DiaSafe
®
Filter / Housing Leak Tests
6. Per the Figure above, disconnect Filter #92’s* tubing from Test Valve #28 to see the valve nozzle. Fluid
output from the nozzle?
*NOTE Water inside Filter #92 does NOT NECESSARILY mean Valve #28 is leaking!.
Yes Valve #28 leaking! Proceed to
page 516, procedure number TMP- 9.0.0.
No Reattach Filter #92 then continue to step #7.
7. Allow three (3) minutes then REPEAT steps #2 through #6 to locate a potential VERY slow leak. If no
leaks are located continue to step #8.
8. If referred to INDUCED POSITIVE PRESSURE TESTS return to the procedure that brought you here as
NOTED.

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals