2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 740
V104SC- 2.0.6 LESS THAN 180 ML COLLECTED
A) Return the blue connector the Bicarbonate
Rinse port.
B) Figure right, REMOVE the
WHITE tubing
from the hydraulic side of the
BICARB
RINSE PORT and direct it into the
cylinder.
C) Expecting flow from the tubing,
return to
RINSE!
D) If flow, allow three (3) pulses THEN measure
ten (10) pulses.
E) Interrupt Rinse by pulling the blue connector
out of the Rinse port!
F) More than 180 ml collected?
Yes More than 180 ml! TWO (2) possible bad components: 1) Restricted Bicarb concentrate
connector OR; 2) Bad Bicarb Rinse port.
No Less than 180 ml! ENSURING the machine was in RINSE prior to checking flow, perform parts
a AND b below:
a) With RINSE now interrupted, reattach the Bicarb rinse port tubing!
b) FIVE (5) possible problems: 1) Figure below, bad wire connection(s) between Valve #108
and Valve #104 OR; 2) Loose ribbon cable at the bibag Interface board (refer to Figure 4C
(page 11)) OR; 3) Bad Valve #108 OR; 4) Bad bibag Interface Board OR; 5) Bad electrical
connection, from Valve #108/Valve #104, inside Distribution Box #2 (see Figure 4D (page
13)).

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals