2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 146
MOTORS- 4.5.0 ! EMPTY = 1 / ISOLATE MOTOR BRUSH CONNECTION
a) UNPLUG the NOTED motor from the distribution board!
b) Using a flashlight, check inside the motor’s distribution board position for ‘white’ corrosion or damaged
male pins. Damage indicates the distribution board may need to be replaced!
c) Figure right, remove the plastic cap from the female
distribution connector. Based on the NOTED motor.
IF the Deaeration Motor: It MUST be connected
between pins 1 and 3 i.e. top and middle
IF the Flow Motor: It MUST be connected between
pins 1 and 5 i.e. top and bottom.
d) Set your CALIBRATED voltmeter to
RESISTANCE ).
e) Touch the meter’s leads together. The meter MUST read less than 1 Ω! Subtract this from the
subsequent measurement.
f) Measure, INSIDE the female connector, BETWEEN the soldered terminals. Less than ten (10) Ω?
Yes Less than 10 Ω! See procedure number MOTORS- 4.6.0 (page 147).
No More than 10 Ω! Bad brushes or motor.
LEFT BLANK INTENTIONALLY

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals