2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 535
AIR- 1.0.1 CHECK FOR AN AIR LEAK AT FILTER #73
Figure below, ENSURING TMP is between 300 and 360 mmHg, for one (1) minute, watch through the
OUTPUT (to machine) side of Dialysate Filter #73. This is Location #1 (L1). Air bubbles seen?
Yes Air seen at L1! See procedure number AIR- 1.0.2 (page 536).
No air seen at L1! Proceed to
page 537, procedure number AIR- 1.0.3.
Figure 74 – Leak Checks (1) / Abbreviated Flow Diagram
Figure 75 – Hydraulics (External & Rear Views)