EasyManuals Logo
Home>Fresenius Medical Care>Medical Equipment>2008T

Fresenius Medical Care 2008T Troubleshoot Manual

Fresenius Medical Care 2008T
768 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #733 background imageLoading...
Page #733 background image
2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 729
SECTION 30 – “V104/V108 STUCK CLOSED” OR V104 ER C = 1
V104/108 Stuck ClosedOR V104 Er C =1 indicates a possible restriction in the Bicarb inlet system, as
sensed by Pressure Sensor #110 (Bic Press)
A) ENSURE the Bicarb inlet tubing is NOT restricted!
B) Figure right, ENSURE the Bicarb connector’s filter is
clean AND the adaptor plug opening is not plugged!
NOTE! If external leaks, “No Water”, or a Flow Error EVER occur address them FIRST!
C) TWO (2) possible scenarios 1) or 2) below:
1) IF occurring during OR after a Cleaning / Disinfection Program (Rinse, Heat Disinfect, etc.)
has stopped: See procedure number V104SC- 1.0.0 (page 730).
2) IF occurring in Dialysis Program: Perform parts a THROUGH c below:
a) If attached to a Bicarb jug ENSURE the Pick Up wand tubing (Figure right)
is NOT kinked!
b) If attached to a Central Bicarb Delivery System (SDS) ENSURE good flow from
the station valve! The valve MUST be left open!
c) Call debug screen 15. Watch V104 Er C (bottom row. 3
rd
column
from left) for three (3) minutes OR until if it EVER = 1, even just
once?
Yes V104 Er C = 1, even just once! Proceed to
page 438, procedure number
CO- 14.0.0.
No V104 Er C REMAINS = ‘0! Whatever was causing V104/108 Stuck Closed” has
cleared for now! Perform parts a THROUGH c below:
a) Call debug screen 0 to watch Flow Error (TOP window) for two (2) minutes. If
EVER = 1 a masked Flow Error is occurring! In this event, DO NOT continue to
troubleshoot “V104/108 Stuck Closed”.
b) If pulling from a Bicarb jug or a Central Bicarb Delivery System (SDS) AND
[Conductivity] increases to more than 13.0 mS Valve #104 is, at least, partially
open!
c) Install a venous dummy chamber in the Level Detector module and attempt to reset
all alarms. Does “V104/108 Stuck Closed” appear?
Yes Valve 104/108 Stuck Closed” appears! Proceed to
page 438, procedure
number CO- 14.0.0.
No Whatever was causing the alarm has cleared.

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals