2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 615
SECTION 22 - “WD: FAIL LONG PULSE” OR “WD: FAIL SHORT PULSE”
Turn the machine OFF for two (2) seconds, then back on. TWO (2) possible scenarios 1) or 2) below:
1) IF “WD: Fail Long Pulse” OR “WD: Fail Short Pulse” does NOT reoccur: Turn the machine off
and on ten (10) times allowing twenty (20) seconds in between as the problem may be intermittent! If
“WD” alarm reoccurs see scenario #2.
2) IF “WD: Fail Long PulseOR “WD: Fail Short Pulse” REOCCURS: Perform parts a THROUGH d
below:
a) Figure below, being
CAREFUL to NOT pull the 24V Power Harness loose, open the card cage!
b) Set your CALIBRATED volt meter DC voltage (V
DC
)
c) Connect the meter’s ground (black) lead to chassis ground (see Figure 2,
page 4)).
d) ENSURING the “WD: Fail Long Pulse” OR “WD: Fail Short Pulse” banner is on the screen,
see procedure number WD- 1.0.0 (page 616).
LEFT BLANK INTENTIONALLY

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals